5 System operation 32
fabian Therapy evolution | SW V5.1.x
Ref: 121003.EN / Date: 26Jan2021
System shutdown
Switching OFF the fabian
Remove patient from the ventilator.
Power OFF any SpO
2
or CO
2
devices
connected to the ventilator.
Press and hold the
start/stop button (1) for 6
seconds.
After 6 seconds, a menu will appear on the
screen stating, Do you want to switch to
standby or shut down the device? .
Press Shut down to power
OFF the device.
After the device has powered OFF, a
notice tone will sound.
Press the Alarm Silence button to confirm system switch OFF.
Hold the button for at least three seconds until the WARNING LED
and audio alarm stops.