fabian Therapy evolution | SW V5.1.x
Ref: 121003.EN / Date: 26Jan2021
12.4.4 PRICO disabling alarms
PRICO is turned OFF and cannot be re-enabled while any of the following alarms are active:
Air supply pressure
Check ET tube
Clean flow sensor
Check SpO
2
sensor
Flow sensor defect
Flow sensor not connected
Input pressure blender
Low SIQ
O
2
sensor calibration error
O
2
value out of range
Oxygen sensor defect
Oxygen sensor used up
Oxygen supply pressure
Patient disconnected
SpO
2
module disconnected
SpO
2
sensor failure
Tube Occlusion
12.4.5 PRICO re-enabling cases
PRICO automatically turns back on in case the following alarms are not active anymore:
Check ET tube
Check SpO
2
sensor
Clean flow sensor
Flow sensor defect
Flow sensor not connected
Low SIQ
O
2
value out of range
Patient disconnected
Tube Occlusion
12.4.6 PRICO errors
Inaccurate measurements can be caused by:
Excessive patient movement
Exposure to excessive illumination, such as surgical lamps (especially those with a xenon
light source), bilirubin lamps, fluorescent lights, infrared heating lamps, or direct sunlight
(correct exposure to excessive illumination by covering the sensor with a dark or opaque
material)
Incorrect sensor application or use
Intravascular dyes such as indocyanine green or methylene blue
Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or
intravascular line.
Significant levels of dysfunctional hemoglobins (
for example
: carboxyhemoglobin or
methemoglobin)
Venous pulsation
Loss of pulse signal can occur in any of the following situations:
The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia
The patient is in cardiac arrest or shock
The sensor is too tight
There is arterial occlusion proximal to the sensor
There is excessive illumination from light sources such as a surgical lamp, a bilirubin lamp, or
direct sunlight

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian and is the answer not in the manual?

ACUTRONIC fabian Specifications

General IconGeneral
TypeNeonatal and Pediatric Ventilator
ManufacturerACUTRONIC Medical Systems AG
Modelfabian
Respiratory Rate1 - 150 bpm
FiO2 Range21 - 100%
HFOV Frequency5 - 20 Hz
Power Supply100-240 V AC, 50/60 Hz
Modes of VentilationSIMV, CPAP, HFOV
Tidal Volume Range2 to 300 ml
PEEP/CPAP0-20 cmH2O
Inspiratory Time0.1 to 3 seconds
Flow Rate2 - 30 L/min
Flow SensorIntegrated
Pressure Limit10-80 cmH2O