73 12 Accessories and options
fabian Therapy evolution | SW V5.1.x
Ref: 121003.EN / Date: 26Jan2021
12.3.4 Alarm delay
You can configure the alarm delay so that transient desaturations do not give an immediate
audible alarm or visual display. You can also select alarm delay settings of 0, 5, 10, and 15
seconds.
Figure 12-5: Alarm delay
12.3.5 SpO
2
averaging time
Selectable averaging time enables clinicians to optimize SpO
2
monitoring for various patient
application areas such as the PACU, NICU, ICU, Telemetry, and Sleep. This reduces the number of
alarms due to real, yet non-clinically actionable rapid desaturations. There are different
averaging times that can be selected (2 to 4, 4 to 6, 8, 10, 12, 14, 16).
Figure 12-6: SpO
2
averaging time

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian and is the answer not in the manual?

ACUTRONIC fabian Specifications

General IconGeneral
TypeNeonatal and Pediatric Ventilator
ManufacturerACUTRONIC Medical Systems AG
Modelfabian
Respiratory Rate1 - 150 bpm
FiO2 Range21 - 100%
HFOV Frequency5 - 20 Hz
Power Supply100-240 V AC, 50/60 Hz
Modes of VentilationSIMV, CPAP, HFOV
Tidal Volume Range2 to 300 ml
PEEP/CPAP0-20 cmH2O
Inspiratory Time0.1 to 3 seconds
Flow Rate2 - 30 L/min
Flow SensorIntegrated
Pressure Limit10-80 cmH2O

Related product manuals