12 Accessories and options 78
fabian Therapy evolution | SW V5.1.x
Ref: 121003.EN / Date: 26Jan2021
12.4.3 PRICO Ventilation modes
PRICO is available on the fabian ventilators in the following modes:
High and Low Flow Oxygen Therapy HFNC (O
2
Therapy)
NIV Bi-levels positive airway pressure (DUOPAP)
NIV Continuous Positive Airway Pressure NCPAP)
The “O
2
Flush can be activated during PRICO, this will deactivate PRICO.
To re-enable PRICO after an O
2
Flush, push the PRICO ON button in
the PRICO menu.
To disable the O
2
Flush function, set the O
2
Flush Time to 0 seconds.
The “Start/Stop button could be used to interrupt mechanical ventilation
for 2 minutes.
In nCPAP, DUOPAP and O
2
Therapy mode ventilation can be interrupted
indefinitely.
In both cases PRICO will be switched OFF.
When PRICO is ON, this is indicated by the PRICO icon in the Status bar and the Green field with
the current FiO
2
labelled PRICO above the O
2
parameter.
Figure 12-12: PRICO “ON

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian and is the answer not in the manual?

ACUTRONIC fabian Specifications

General IconGeneral
TypeNeonatal and Pediatric Ventilator
ManufacturerACUTRONIC Medical Systems AG
Modelfabian
Respiratory Rate1 - 150 bpm
FiO2 Range21 - 100%
HFOV Frequency5 - 20 Hz
Power Supply100-240 V AC, 50/60 Hz
Modes of VentilationSIMV, CPAP, HFOV
Tidal Volume Range2 to 300 ml
PEEP/CPAP0-20 cmH2O
Inspiratory Time0.1 to 3 seconds
Flow Rate2 - 30 L/min
Flow SensorIntegrated
Pressure Limit10-80 cmH2O