61 8 Battery operation
fabian Therapy evolution | SW V5.1.x
Ref: 121003.EN / Date: 26Jan2021
Power Failure
In the event of a Power Failure the device
will automatically switch to Battery Power
without interruption.
A visual and audible message is triggered
“Power failure
This message can be confirmed
with the Alarm Silence button.
The device will continuously monitor the remaining battery run time.
Additional messages will be triggered at a remaining Run Time of:
15 minutes
30 minutes
60 minutes, only if the remaining run time is less than 60 minutes during a Power
Failure
If the remaining Run Time is less than 1
minute a warning will appear prompting to
immediately switch to a Primary Power
Source.
After 60 minutes on Battery power, the
battery is drained and fabian will power
OFF.
Depleted Battery!
Connect to Mains
Charge battery for approx. 5 hours

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian and is the answer not in the manual?

ACUTRONIC fabian Specifications

General IconGeneral
TypeNeonatal and Pediatric Ventilator
ManufacturerACUTRONIC Medical Systems AG
Modelfabian
Respiratory Rate1 - 150 bpm
FiO2 Range21 - 100%
HFOV Frequency5 - 20 Hz
Power Supply100-240 V AC, 50/60 Hz
Modes of VentilationSIMV, CPAP, HFOV
Tidal Volume Range2 to 300 ml
PEEP/CPAP0-20 cmH2O
Inspiratory Time0.1 to 3 seconds
Flow Rate2 - 30 L/min
Flow SensorIntegrated
Pressure Limit10-80 cmH2O