6146/6156 DC Voltage/Current Generators Operation Manual
7.2.2 Outlined Calibration Procedure
7-5
7.2.2 Outlined Calibration Procedure
The calibration procedure is outlined below. Always execute DAC calibration before voltage/current
source calibration.
Figure 7-2 Outlined Calibration Procedure
Refer to Section 7.1.2, “Safety Precautions” (3).
Refer to Section 7.2.3, “DAC Calibration Procedure (6156 Only).”
NOTE: Always execute voltage/current source calibration after
DAC calibration.
Refer to Section 7.2.4, “Calibration Using Remote Commands,”
or Section 7.2.5, “Calibration by Manual Operation.”
Calibration
Warm-up
Power ON
DAC calibration
Voltage/current calibration
Voltage/current limit calibration
Calibration end