6146/6156 DC Voltage/Current Generators Operation Manual
7.2.4 Calibration Using Remote Commands
7-9
Figure 7-6 Remote Calibration Procedure (Voltage Limiter)
Refer to Table 7-2, "Calibration Tolerance Ranges."
Voltage limiter calibration
OPR
Calibration mode operation
Polarity setting
Positive
polarity?
No
Yes
XF2
VL mode
XS0
Positive (VL-HI)
XS1
Negative (VL-LO)
XN0
ZERO point calibration (
±1V)
XM1
DMM input mode
[ZERO]
DMM data
acquisition
XDT DMM data
DMM data input *Out-of-range setting results in a parameter error.
ex. “XDT1.02”
XN1
FS calibration (
±30V)
XM1
DMM input mode
[FULL]
DMM data
acquisition
XDT DMM data
DMM data input *Out-of-range setting results in a parameter error.
ex. “XDT30.32”
XN0
ZERO point calibration (
±1V)
[ZERO]
DMM data
acquisition
Calibration result data
save/display
XN1
FS calibration (
±30V)
[FULL]
DMM data
acquisition
Calibration result data
save/display
Within the
tolerance range
No
Yes
No
Yes
Complete?
SBY
Calibration mode standby