6146/6156 DC Voltage/Current Generators Operation Manual
7.2.4 Calibration Using Remote Commands
7-8
Figure 7-5 Remote Calibration Procedure (Voltage Source)
Refer to Table 7-2, "Calibration Tolerance Ranges."
XR2
Voltage source
calibration
OPR
Calibration mode operation
Range setting
30 mV ?
No
Yes
300 mV ?
No
Yes
3 V ?
No
Yes
XR3
300 mV range
XR4
3 V range
XR5
30 V range
Polarity setting
Positive
polarity?
No
Yes
XS0
Positive
XS1
Negative
XN0
ZERO point calibration
XM1
DMM input mode
[ZERO]
DMM data
acquisition
XDT DMM data
DMM data input *Out-of-range setting results in a parameter error.
ex. “XDT-2. 0e-6”
XN1
FS calibration
XM1
DMM input mode
[FULL]
DMM data
acquisition
XDT DMM data
DMM data input *Out-of-range setting results in a parameter error.
ex. “XDT30. 32e-3”
XN0
ZERO point calibration
[ZERO]
DMM data
acquisition
XN1
FS calibration
Calibration result data
save/display
[FULL]
DMM data
acquisition
Calibration result data
save/display
Within the
tolerance range?
No
Yes
No
Yes
Complete?
SBY
Calibration mode standby
XF0
VS mode
30 mV range

Table of Contents