6146/6156 DC Voltage/Current Generators Operation Manual
7.2.4 Calibration Using Remote Commands
7-11
Figure 7-8 Remote Calibration Procedure (Current Limiter)
Refer to Table 7-2, "Calibration Tolerance
Ranges."
Current limiter
calibration
OPR
Calibration mode operation
Polarity setting
Standard resistance
1
Ω
connection
No
Yes
XF3
IL mode
Positive polarity?
XS0
Positive (IL-HI)
XS1
Negative (IL-LO)
XN0
ZERO point calibration (minimum value)
XM1
DMM input mode
[ZERO]
DMM data acquisition
Current conversion
Converted current value (A) = DMM measured voltage (V) /
Standard resistance for calibration (
Ω
)
ex. 1.02mV / 1Ω = 1.02mA
XDT converted data
Converted current data input *Out-of-range setting results
in a parameter error. ex. “XDT1.02e-3”
XN1
FS calibration
XM1
DMM input mode
[FULL]
DMM data acquisition
Current conversion
Converted current value (A) = DMM measured voltage
(V) / Standard resistance for calibration (
Ω
)
ex. 200.32mV / 1Ω = 200.32mA
XDT converted data
Converted current data input *Out-of-range setting results in
a parameter error. ex. “XDT200.32e-3”
XN0
ZERO point calibration
[ZERO]
DMM data acquisition
Current conversion
Calibration result data
save/display
XN1
FS calibration
[FULL]
DMM data acquisition
Current conversion
Within the
tolerance range
Calibration result data
save/display
No
Yes
No
Yes
Complete?
SBY
Calibration mode standby