6146/6156 DC Voltage/Current Generators Operation Manual
A.2 Error Message List
A-5
Self test 251 Voltage source 30 V-FS test -30 V Full Check
260 Current source 3 mA+ZERO test +3 mA Zero Check
261 Current source 3 mA +FS test +3 mA Full Check
262 Current source 3 mA -ZERO test -3 mA Zero Check
263 Current source 3 mA -FS test -3 mA Full Check
264 Current source 30 mA+ZERO test +30 mA Zero Check
265 Current source 30 mA +FS test +30 mA Full Check
266 Current source 30 mA -ZERO test -30 mA Zero Check
267 Current source 30 mA -FS test -30 mA Full Check
268 Current source 200 mA+ZERO test +200 mA Zero Check
269 Current source 200 mA +FS test +200 mA Full Check
270 Current source 200 mA -ZERO test -200 mA Zero Check
271 Current source 200 mA -FS test -200 mA Full Check
280 CMV test Cmv Test
290 Relay life check (range) Rly Limited 1
291 Relay life check (Operate) Rly Limited 2
Hardware error 401 Fan stop alarm Fan Stop Alarm
402 Overheat alarm Over Heat Alarm
403 Source unit circuit malfunction Source Unit Error
404 Overload alarm Over Load Alarm
Source condition
error
601 Out-of-range thermal electromotive force output Out of range for TC
831 Output disabled by interlock Interlock
Command error -102 Command syntax error Cmd Syntax
-113 Unsupported command Undefined header
-131 Suffix error (e.g. SOV1A) Invalid suffix
-200 Command currently not executable Execution error
-222 Out-of-range parameter setting Data out of range
System -313 Calibration data is lost. Calibration memory lost
-314 Data saved by the *SAV command is lost. Save/recall memory lost
-315 Saved setting data is lost. Configuration memory lost
-350 Overflowed error queue Queue overflow
-330 Self test error Self-test failed
Calibration 500 Calibration data error CAL data
501 Unexecuted ZERO point calibration before ZERO
USB
communication
150 USB communication error Illegal packet received
Table A-2 Error Message List (2 of 2)
Classification
Error
code
Description ERR? command response