6146/6156 DC Voltage/Current Generators Operation Manual
A.3.1 GPIB/USB Remote Execution Time (Typical Value)
A-7
2. Source value data read time
Voltage source* SOV+32.0000E+0 (No range change) PRGM MODE: DC
Output status: Operate
Other: default
5.4 3.7
(Range change) 87 85
Current source* SOI+220.000E-3 (No range change) 6.1 3.6
(Range change) 37 35
Thermal electromo-
tive force source*
TMP+0100.0E+0 36 35
TMP-0100.0E+0 61 59
Voltage limit
value*
LMV-001.0E+0, +001.0E+0 6.9 4.3
Current limit
value*
LMI-005.0E-3, +005.0E-3 6.8 4.6
Time parameter* SP10.00E+0 3.5 3.3
Source mode MD0, AMD2, AMD4 PRGM MODE: DC
Output status: Standby
Other: default
17 17
Memory data set-
ting*
N0, SOV+320.000E-3V 6.7 4.2
*The command with <data> is different in processing time according to the data length.
Item Command Condition
GPIB
Unit [ms]
USB
Unit [ms]
Query command SOV? Default 3.7 2.8
SOI? 3.7 2.8
TMP? 5.8 5.4
D? 3.2 2.6
Reading not by a query command 2.1 1.9
Item Command Condition
GPIB
Unit [ms]
USB
Unit [ms]

Table of Contents