6243/44 DC Voltage Current Source/Monitor Operation Manual
6.4.2 Data Output Format in the TR6143 Mode (Talker)
6-17
•Flag
V/I: V/I = 0; Indicates the current measurement data
V/I = 1; Indicates the voltage measurement data
Range code
D7 D6 D5 Description
0 0 0 Current or voltage measurement data
0 0 1 Data when the comparison operation result is GO
0 1 0 Data when the comparison operation result is LO
0 1 1 Data when the comparison operation result is HI
1 0 0 Data when the limiter (LMT) performs.
1 0 1 Data output when an oscillation is detected (OSC)
1 1 0 Over range data
V/I Range code
Measurement data Range
D4 D3 D2 D1 D0
10
0
0
0
1
1
1
1
0
0
1
1
0
1
0
1
Voltage
320 mV
3.2 V
32 V/20 V
110 V
00
0
0
0
0
1
1
0
1
1
1
1
0
0
1
0
0
1
1
0
0
1
0
1
0
1
0
1
Current
32 A
320 A
3.2 mA
32 mA
320 mA
2 A/3.2 A
10 A