6243/44 DC Voltage Current Source/Monitor Operation Manual
6.4.2 Data Output Format in the TR6143 Mode (Talker)
6-18
•Data
• Complement-on-two data
• Depending on the Resolution setting, the byte length changes as follows:
• Examples of data when Resolution is 4-1/2 digits
(4) Measurement data number output
When measurement data number output is specified by OM3, the measurement data number is output
in the following format.
(a) Header (ON/OFF possible by S5/S4 command)
(b) Data number
(c) Block delimiter
• Data number (16-bit binary data output in the order from high order to low order)
Resolution Byte Bit Data range Default
4-1/2 digits 2 Byte b15 to b0 -32000 to +32000
5-1/2 digits 3 Byte b23 to b0 -320000 to +320000
b15 b14 b13 b12 b11 b10 b9 b8 b7 b6 b5 b4 b3 b2 b1 b0 Data
0
0
0
1
1
1
0
0
1
0
1
0
0
1
0
1
0
0
1
0
1
0
0
1
0
1
0
0
1
0
0
0
0
1
1
1
0
0
1
1
0
0
0
1
0
0
0
0
1
0
0
0
0
1
0
0
0
0
1
0
0
0
0
1
0
0
0
0
1
0
0
0
0
1
0
0
1
0
1
0
+32000
+1
0
-1
-32000
DC
(a) (b) (c)
b15 b14 b13 b12 b11 b10 b9 b8 b7 b6 b5 b4 b3 b2 b1 b0 Decimal
number
The number
of data
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
1
0
0
1
0
0
0
0
0
0
0
1
1
0
0
1
1
0
0
1
1
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
0
1
0
0
0
1
0
1
0
1
0, 0
0, 1
11, 128
19, 135
0
1
2944
4999