Corner area of detector not exposed
Details A small area in the corner of the detector has not been ex-
posed.
Cause When making an exposure with small source-image distance
(e.g. 1 m) and with the collimator in rotated position (e.g.
45°), the X-ray field does not reach to the outer corners of the
collimated area.
Solution Increase the source-image distance.
DR 100s | Problem solving | 155
0411D EN 20220627 1040