AGFA DR 100s - X-ray Specifications (40 kW Configuration)

AGFA DR 100s
198 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
X-ray group (40 kW configuration)
Table 31: X-ray specifications
Nominal power (IEC 60601-1) 40 kW @ 100 kV, 400 mA
kV values 40 - 150 kV at steps of 1 kV
Inverter frequency in high voltage 40 kHz
mA values
70 - 400 mA
(automatically associated to kV)
mAs values
0.1 - 500 mAs
(with increases of 12.5%)
Exposure times
0.001 - 3 s (DR mode)
0.001 - 10 s (CR mode)
(according to mAs)
Table 32: Accuracy (IEC 60601-2-54)
kV ±5%
mA ±20%
mAs ±(10% + 0.2 mAs)
Time ±(10% + 0.1 ms)
Table 33: Equivalent filtration @75 kV
Filtration 1.5 mm Al
Collimator 2 mm Al
Total filtration 3.5 mm Al
Additional filtration of DAP meter (optional) 0.3 mm Al
Total filtration with DAP meter (optional) 3.8 mm Al
DR 100s | Product Information | 179
0411D EN 20220627 1040