Table 34: Maximum settable mAs
focus small small large large
exposure
time
1 s 3 s 1 s 3 s
minimum
mAs
0.1 0.1 0.63 0.63
kV maximum
mAs
maximum
mAs
maximum
mAs
maximum
mAs
40 200 280 320 500
50 200 280 320 500
60 160 250 320 500
70 160 250 320 500
80 125 180 250 360
90 125 180 250 360
100 125 180 200 280
110 100 160 200 280
120 100 160 160 250
130 80 125 160 250
140 80 125 125 250
150 80 125 125 250
Table 35: Generator
Model HF1 G/29
Maximum power (100 kV - 400 mA) 40 kW
Maximum voltage to the tube 150 kV
Maximum current to the tube 450 mA
Ripple to the maximum power < 2%
Rise time to the max. power < 1 ms
Thermal safety 60 °C ± 5° C
Oil volume compensation lung 0.20 dm
3
180 | DR 100s | Product Information
0411D EN 20220627 1040