Using the wired exposure button
Completely extend the cable of the exposure button and keep as far as possible
from the radiation source. The cable of the exposure button can be extended
up to 4 m. Stretching it further can damage the cable.
1. Press and hold the exposure button down to the first pressure point.
The X-ray tube is prepared for performing an exposure.
To start the exposure directly after preparation of the X-ray tube has
ended, immediately press the exposure button down fully.
2. Press the exposure button down fully and keep it pressed until the
exposure has ended.
When the X-ray exposure is finished successfully, three beeps are heard.
WARNING:
During exposure ionizing radiation is emitted by the X-ray
system. To indicate the presence of ionizing radiation, the
radiation indicator on the control console lights up.