Connecting USB devices
WARNING:
Only USB devices powered via the USB cable may be connected
to a USB port of the PC of the mobile X-ray system. The use of
USB devices powered by an AC/DC power supply is strictly
forbidden.
CAUTION:
The USB device must have been certified acc. to CISPR11 or
CISPR22 (or EN 55011 or EN 55022 accordingly), class A
(minimum).
CAUTION:
If the USB device causes radio interference or disrupts the
operation of nearby equipment, it may be necessary re-orient or
relocate the device or shielding the location.
DR 100s | Product Information | 163
0411D EN 20220627 1040