Equipotential earth connection
The device is supplied with an equipotential earth connection point.
The device can only be used in areas that comply with local electrical safety
norms and in environments suitable for medical activities.
An additional equipotential earth connection is provided because the
equipment is movable and the reliability of the main equipotential earth
connection point can be insufficient.
It is only possible to use this equipment in rooms that comply with IEC norm
requirements.
60 | DR 100s | Introduction
0411D EN 20220627 1040