6 V in ISM bands
from 150 kHz to
80 MHz
6 V in ISM bands
from 150 kHz to
80 MHz
used no closer than
30 cm to any part of
the equipment, in-
cluding cables speci-
fied by manufacturer.
Voltage dips,
short inter-
ruptions and
voltage varia-
tions on pow-
er supply in-
put lines.
IEC
61000-4-11
0% U
T
for 0.5 cy-
cle
0% U
T
for 1 cycle
70% U
T
for 25
cycles
0% for 5 s
0% U
T
for 0.5 cy-
cle
0% U
T
for 1 cycle
70% U
T
for 25
cycles
0% for 5 s
The quality of the
voltage supply should
correspond to that of
a typical commercial
or clinical environ-
ment. If the user
wants the device to
work continuously,
even when the energy
supply is interrupted,
it is recommended to
use an energy supply
free of interruptions
or a battery.
Magnetic field
at the supply
frequency
(50/60 Hz)
IEC 61000-4-8
30 A/m 30 A/m Magnetic field at the
network frequency
should correspond to
the typical values as
they are in a commer-
cial and clinical envi-
ronment.
Note: U
T
is the alternating current in the network before the application
of the test level.
Topics:
Immunity to RF wireless communication equipment
Precautions on EMC
Cables, transducers and accessories
For U.S.A.
194 | DR 100s | Product Information
0411D EN 20220627 1040