1290 Infinity Quaternary Pump User Manual 277
LAN Configuration
13
Automatic configuration with Bootp
Adding each instrument to the network using BootP
1 Follow Start > All Programs > Agilent BootP Service and select Edit BootP
Settings. The BootP Settings screen appears.
2 Uncheck the Do you want to log BootP requests? once all instruments have
been added.
The Do you want to log BootP requests? box must be unchecked when you
have finished configuring instruments; otherwise, the log file will
quickly fill up disk space.
3 Click Edit BootP Addresses... The Edit BootP Addresses screen appears.
4 Click Add... The Add BootP Entry screen appears.
Figure 47 Enable BootP logging
5 Make these entries for the instrument:
• MAC address
• Host name, Enter a Hostname of your choice.
The Host Name must begin with "alpha" characters (i.e. LC1260)
• IP address
• Comment (optional)