278 1290 Infinity Quaternary Pump User Manual
13 LAN Configuration
Automatic configuration with Bootp
Subnet mask
Gateway address (optional)
The configuration information entered is saved in the Tab File.
6 Click OK.
7 Leave Edit BootP Addresses by pressing Close.
8 Exit BootP Settings by pressing OK.
9 After each modification of the BootP settings (i.e. EditBootPSettings) a
stop or start of the BootP service is required for the BootP service to
accept changes. See “Stopping the Agilent BootP Service” on page 279
or “Restarting the Agilent BootP Service” on page 280.
10 Power cycle the Instrument.
OR
If you changed the IP address, power cycle the instrument for the
changes to take effect.
11 Use the PING utility to verify connectivity by opening a command
window and typing:
Ping 192.168.254.11 for example.
The Tab File is located at
C:\Documents and Settings\All Users\Application Data\Agilent\BootP\TabFile