110 Agilent Restricted Agilent 5110/5100 ICP-OES Service Manual
4 Removal/ Installation, Replacement and Adjustment
Gas Filter
1 To remove Gas filter slide out from spring holder.
2 Press quick connect rim in direction of red arrow and pull tubing out (yellow) for both top and
bottom fitting.
3 When replacing filter note direction of flow.