Agilent 5110/5100 ICP-OES Service Manual Agilent Restricted 121
4 Removal/ Installation, Replacement and Adjustment
Pre Optics System
3 Remove 4 torx screws from motor.
4 Disconnect motor and remove from casting.
Mirror radial M2
1 Remove pre optics side cover.
2 Remove torx screw holding mirror to casting and remove mirror assembly (including
bracket) from casting.
3 Reinstall replacement mirror assembly which is factory pre aligned. Locate mirror bracket
hole on lower dowel pin and hold bracket against upper dowel pin before tightening torx
screw.
4 Create a Manganese (Mn) method (refer Diagnostics section) and run a time scan. If required
make minimal adjustments to Radial mirror M2 to maximize intensity in the horizontal plain.