268 Agilent Restricted Agilent 5110/5100 ICP-OES Service Manual
10 Installation Guide (5110 ICP-OES)
Install ICP-OES System
Install ICP-OES System
Following is an outline of the order of tasks you should perform during installation. A detailed
explanation of procedures follows later in the chapter:
1 Connect power to the instrument and turn on mains switches. The instrument firmware will
turn on the polychromator heaters automatically and it may take up to 4 hours to stabilize at
+35°C.
2 Connect the argon supply to the instrument.
3 Connect the water cooler to the instrument.
4 Check extraction system flow and connect to instrument.
5 Ensure the Argon supply has been running for at least 20 minutes before switching on the
water cooler.
6 Install the sample introduction components.
7 Install the Agilent ICP Expert 7 software.
8 Install the LAN cable between Instrument and PC.
9 Configure the IP Address.
10 Complete the Agilent software registration requirements.
11 Switch on the instrument and check that the instrument initializes correctly.
12 Check the Argon Ratio and adjust if required.
13 Perform a Wavelength Calibration (including a dark current scan).
14 Run the performance tests to check the instrument:
Resolution
Signal to Background Ratio (SBR)
Calibration accuracy
Precision
15 Re-check the Argon Ratio with the polychromator stabilized at +35°C. Ensure it is the same
as the factory setting and adjust if required.
Section Contents:
Hardware installation 268
Mains connection 270
Gas connection 270
Water chiller connection 271
Water chiller connection to the spectrometer 272
Software installation 273
Hardware installation
The 5110 instrument is packaged in a timber crate. Inspect for external damage with customer
before unpacking.

Table of Contents

Other manuals for Agilent Technologies 5100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5100 and is the answer not in the manual?

Agilent Technologies 5100 Specifications

General IconGeneral
BrandAgilent Technologies
Model5100
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals