128 Agilent Restricted Agilent 5110/5100 ICP-OES Service Manual
4 Removal/ Installation, Replacement and Adjustment
Polychromator
Removal of the polychromator outer box top cover provides access to:
Shutter assembly
Entrance slit plug
Grating and prism casting assy
Back lighting fixture access cover
Polychromator and camera purge gas lines
Camera rotation adjustment screws access plugs
Full illumination LED fibre
1 Remove 4 pozidrive 2 point end cover access screws.
2 Remove 13 pozidrive 2 point screws that secure the top cover to the side walls and remove
the polychromator heater box top cover.
Camera rotation
Camera rotation
Camera access
Slit access Screw
DLA/Focus/W/L
adjustment
Back lighting jig
location
Grating prism
assembly