EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>5100

Agilent Technologies 5100 Service Manual

Agilent Technologies 5100
312 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #150 background imageLoading...
Page #150 background image
150 Agilent Restricted Agilent 5110/5100 ICP-OES Service Manual
4 Removal/ Installation, Replacement and Adjustment
Argon ratio adjustment
NOTE
Location adjustments should be complete before attempting focus adjustments.
The argon ratio adjustment is the final adjustment in any alignment. The argon ratio is also set
at installation.
NOTE
The Argon ratio value is stored in Instrument Memory (See Figure 1 below) (S/W Versions 7.3
and above) and can be found on the Status Page under “More Hardware information”. Only
change this value if the Camera is replaced. The value is changed by entering Diagnostics using
the password and clicking on “Edit” on the “Hardware information Page”, editing value, then “Set”
to save data.
Figure 17.
1 Remove the polychromator access cover from the left side of the instrument.
2 Access the mirror adjustment screws by lowering the insulation flap on the polychromator
outer box insulation.

Table of Contents

Other manuals for Agilent Technologies 5100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5100 and is the answer not in the manual?

Agilent Technologies 5100 Specifications

General IconGeneral
BrandAgilent Technologies
Model5100
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals