264 Agilent Restricted Agilent 5110/5100 ICP-OES Service Manual
9 iTroubleshoot/Parts Predict
Problem Solving Table
Water Cooling Water Cooling Ready LED Red Water Temp out of spec Replace Temp sensor
Water Cooling Water Cooling Ready LED Red Water Temp out of spec Replace Instrument Control PCA G8010-65200
Water Cooling Water Temperature high Chiller Air inlet obstructed Clean filter - Ensure air flow
around chiller
Wavelength calibration Fails calibration at high and low
wavelengths
Argon Ratio incorrect Adjust Argon Ratio
Wavelength calibration Fails calibration at high and low
wavelengths
Calibration solution Use Fresh calibration solution 6610030100
Wavelength calibration Fails calibration at high and low
wavelengths
Moisture on Detector Lens Purge gas contaminated. Verify
using Polyalign. Should be no
peaks with Shutter closed.
Switch off Chiller and wait 30
minutes and then retry.
Wavelength calibration Fails calibration at high and low
wavelengths
Optimize Pre optics Optimize Pre Optics using Mn
Line
Wavelength calibration Fails calibration at high and low
wavelengths
Realign Polchromator Realign Polychromator including
Focus
Wavelength calibration Fails calibration at high and low
wavelengths
Replace Camera Replace Camera as a last resort G8014-60100
Wavelength calibration Fails Calibration at low
wavelengths only
Fails low wavelengths Increase purge time, Ensure
Boost purge, cone and snout
purge are on.
Wavelength calibration Fails Calibration at low
wavelengths only
Windows Contaminated Clean or replace Axial or Radial
Windows
G8010-68014,
G8010-68015
Wavelength calibration Fails Calibration at low
wavelengths only
Holes in Optics not sealed Check for grommet sealing
around Optics and no screws
missing
Table 1
Symptom Area Problem Problem Description Solution Part Number(s)

Table of Contents

Other manuals for Agilent Technologies 5100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 5100 and is the answer not in the manual?

Agilent Technologies 5100 Specifications

General IconGeneral
BrandAgilent Technologies
Model5100
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals