Agilent 5110/5100 ICP-OES Service Manual Agilent Restricted 289
10 Installation Guide (5110 ICP-OES)
Analytical Tests
Precision
The precision for the 10 replicates of sample readings on the multi element standard should be
less than the following:
Table 5 Agilent 5110 ICP-OES Instruments
Element Specification
Al167, As, Pb & Se 2.6 %RSD
All other lines 1.5 %RSD