Maintenance
114 Agilent 55B AA Spectrometer User’s Guide
To install the spray chamber/nebulizer assembly:
1 Place the spray chamber/nebulizer assembly in position in the sample
compartment so that the fuel and oxidant outlet nipples on the burner
adjuster. (see Figure 74 below) are engaged in the fuel and oxidant ports in the
side of the nebulizer block.
Figure 74.
Fuel and oxidant outlet nipples
Figure 75.
Retaining screws
2 Push the assembly firmly sideways. Tighten the knurled retaining screws
alternately, one turn each at a time, so that the assembly is pulled up evenly
into place (see Figure 75 above). Tighten both screws firmly with the fingers
only.
3 Place a suitable wide-necked waste vessel in a convenient position at floor
level.
The waste vessel must be made of a non-glass material which will not
shatter in the event of a flashback or fire.
Locate your waste vessel in an open well-ventilated area where you can
see it.
Never locate the waste vessel in a confined space.
Use small stable vessels and empty them frequently.
WARNING
Fire Hazard.
Do not accumulate large volumes of flammable liquid.
Dispose of flammable liquids in accordance with local regulations.
4 Attach a convenient length of 9 mm ID tubing to the lower outlet of the liquid
trap.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 55B AA and is the answer not in the manual?

Agilent Technologies 55B AA Specifications

General IconGeneral
BrandAgilent Technologies
Model55B AA
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals