Part 6. Calibrating your capillary column (optional) Columns and Traps
Column calibration procedures
8. To recalibrate or change calibration mode, press [Mode/Type] to see the
column calibration mode menu.
9. Scroll to Length or Diameter and press [Enter]. The following menu
appears:
10. Scroll to Measured flow and enter the corrected column flow rate (in
mLmin) from the run performed above.
11. When you press [Enter], the GC will estimate the column length or diameter
based on the elution time input and will now use that data for all calculations.
The * indicates the current mode
Available calibration modes
Actual column flow rate, mL/min (input)
Elution time, min (input)
Calc length: current estimated column
length based on input
Calc diameter: current estimated colum
n
ID based in input