223
Analytical Sequences
Defining a sequence
4. Press [Mode/Type] and select one of the three injector types.
5. Supply the rest of the subsequence parameters. If you are using both injec-
tors, there will be two sets of parameters.
#Injections/vial—the number of repeat runs from each vial. Enter
0 if no samples are to be injected.
Samples—the range (first-last) of sample vials to be analyzed.
6. If this is the Priority Sequence, set Use priority to On.
7. Proceed to the next subsequence or to the Post Sequence.
Procedure: Creating a valve subsequence
If your GC is equipped with a gas sampling valve and an (optional) multiposition
valve, a valve subsequence can be created.
1. Press [Seq] to open the sequence control table.
2. Scroll to a subsequence Method # line. If this is the Priority Sequence,
the line is labeled Priority meth #.
3. Enter a method number. Use 0 for the currently active method, 1 to 5 for the
stored methods, or Off to end the sequence.
The active method, 0, will change during the sequence if the subsequences
use stored methods. Therefore, method 0 should be chosen for the Priority
Sequence only if all subsequences use method 0.
4. Press [Mode/Type] and select Valve.
Sampler sequence
Press [Mode/Type] to select the type.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 GC and is the answer not in the manual?

Agilent Technologies 6890 GC Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 GC
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals