Analytical Sequences
Defining a sequence
224
Figure 40 Valve subsequences
5. Enter the valve sequence parameters (the first three appear only if a multi-
position valve is configured):
#inj/position number of injections at each position, (0-99)
Position rng first-last valve positions to sample, (1-32)
Times thru range number of times to repeat the range, (1-99)
# injections number of injections for each sample
Procedure: Setting the Post Sequence events
1. Scroll to the Post Sequence title section.
2. Method # is the method to be loaded and run once at the end of a sequence.
Enter 1 through 5 for stored methods. If there is no method to be loaded,
enter 0.
3. Repeat sequence—On keeps repeating the sequence. This function is
useful for valve sequences. Off halts the sequence at the end. Turn Repeat
sequence On or Off.
With multiposition valve
Without multiposition valve