191
Valve Control
The valve box
Valve temperature programming
Most valve applications are isothermal; however, you can define three
temperature ramps if desired. Press [Aux #], then [1] or [2]. Program this ramp
the same as an oven ramp. Refer to Making a temperature-programmed run
on page 88 for more information.
Configuring an Aux thermal zone
To configure a thermal Aux zone (1 or 2), press [Config], then [Aux #]. Press
[Mode/Type], then select the type of device to be controlled by the zone and
press [Enter].