EasyManuals Logo

Agilent Technologies 6890 GC User Manual

Agilent Technologies 6890 GC
286 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #130 background imageLoading...
Page #130 background image
Columns and Traps Part 1. Capillary columns
Procedure: Installing capillary columns in the FPD
118
Procedure: Installing capillary columns in the FPD
The FPD uses an adaptable fitting that can use both packed and capillary
columns. If your adaptable fitting does not have a capillary adapter installed,
begin with step 1. If the capillary adapter is already installed in your adaptable
fitting, begin with step 5.
The FPD uses a special adapter for capillary columns. The FPD Capillary
Adapter, part number 19256-80570, allows fused silica columns as large as
530 mm ID to be run right to the base of the FPD flame, minimizing sample tailing
or loss of chemically active sites.
Materials required
Column nut and ferrule
FPD Capillary column adapter
1/4-inch nut and ferrule
Column cutter
1/4-inch wrench
9/16-inch wrench
Metric ruler
Typewriter correction fluid
19256-80590
FPD 1/8-inch adapter
19256-80570
FPD Capillary Adapter

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 GC and is the answer not in the manual?

Agilent Technologies 6890 GC Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 GC
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals