Table 24 Bad Mainboard and Fatal Error Messages
Popup message Comments
BAD MAINBOARD
Main FPGA Failure
Contact your Agilent service representative.
Static RAM Failure
Boot ROM Checksum
ROM #2 Checksum Either EEPROM 2 or 3 is malfunctioning. Contact your Agilent
service representative.
ROM #3 Checksum
Incorrect ROM #2 EEPROM 2 or 3 is installed in the wrong position. Change the
position the EEPROM is installed in; contact your Agilent service
representative for more information.
Incorrect ROM #3
ROM #2 wrong version The version of either EEPROM 2 or 3 is different than that of
the other EEPROMs. Contact your Agilent service
representative.
ROM #3 wrong version
DMA FPGA Failure
Contact your Agilent service representative.
DRAM Failure
FATAL ERROR
Exception Vector
Contact your Agilent service representative.
Bus Error
Address Error
Illegal Instruction
Divide by Zero
No 512Hz Interrupt