245
Part 1. Not Ready messages Messages
Other not ready messages
Power on in progress
This message appears when:
• Power is restored after a power failure during a run or while the oven was
turned on and the GC was not performing a run.
• Power is turned on again after a user turned it off while the oven was turned
on.
The GC heats all the other thermal zones and then heats the oven. When the
oven temperature reaches the setpoint for equilibration time, the GC becomes
ready.
If the power failure occurred during a run, upon power restoration the GC heats
all the thermal zones and the oven and automatically performs a blank run. When
the blank run is completed, the GC becomes ready.