Messages Part 3. Warning messages
254
Table 27, continued
Warning no. Status message Popup message Run log entry Comments
128 F inl flow cal fail Front inlet flow sensor
auto zero calib failed.
Contact Agilent service.
129 B inl flow cal fail Back inlet flow sensor
auto zero calib failed.
Contact Agilent service.
130 Aux 1 cryo disabled Aux 1 & front inlet on
same cryo valve drive:
aux1 disabled
Reconfigure aux or inlet
cryo drive.
131 Aux 2 cryo disabled Aux 2 & back inlet on
same cryo valve drive:
aux2 disabled
Reconfigure Aux or inlet
cryo drive.
132 Chgd Col 1 Init time to
###.## ; avoids Sampling
End problem
For Volatiles interface, a
setpoint conflicted with
the Sampling End time
parameter. Check your
method. See the Inlets
volume for more
information.
133 Chgd Col 2 Init time to
###.## ; avoids Sampling
End problem
134 Chgd FI Saver time to
###.## ; avoids Sampling
End problem
135 Chgd BI Saver time to
###.## ; avoids Sampling
End problem
136 Chgd FI Purge time to
###.## ; avoids Sampling
End problem
137 Chgd BI Purge time to
###.## ; avoids Sampling
End problem

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6890 GC and is the answer not in the manual?

Agilent Technologies 6890 GC Specifications

General IconGeneral
BrandAgilent Technologies
Model6890 GC
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals