Messages Part 4. Fault messages
260
Table 28, continued
Fault no. Status message Popup message Run log Comments
255 F det invalid type Front detector: invalid det
module
Not ready:
F det invalid type
256 B det invalid type Back detector: invalid det
module
Not ready:
B det invalid type
257 F inlet invalid type Front inlet: invalid inlet
module
Not ready:
F inlet invalid type
258 B inlet invalid type Back inlet: invalid inlet
module
Not ready:
B inlet invalid type
259 F det type mismatch Front detector: det board
not the same as module
Not ready:
F det type mismatch
If you installed a new
detector, check that the new
detector’s electronics board
and module are installed in
the proper locations.
260 B det type mismatch Back detector: det board
not the same as module
Not ready:
B det type mismatch
261 MIO board defective Host communications:
MIO board defective
Not ready:
MIO board defective
262 RS232 defective Host communications:
RS232 defective
Not ready:
RS232 defective
263 GPIB defective Host communications:
GPIB defective
Not ready:
GPIB defective
264 Sampler RS232 defect Sampler communications:
RS232 defective
Not ready:
Sampler RS232 defect
265 F inlet invalid pid Front inlet: invalid pids
266 B inlet invalid pid Back inlet: invalid pids
267 F det invalid pid Front detector:
invalid pids
268 B det invalid pid Back detector:
invalid pids
269 Pneu aux invalid pid Pneu aux module:
invalid pids
270 F inlet bad cksum Front inlet:
invalid module checksum