11-44 Programmer’s Guide
SCPI Command Summary
SCPI Device Command Summary
Table 11-19 SENSe (7 of 16)
1. For use with multiport test sets only.
2. Numeric parameters may include an appropriate suffix; if no suffix is included, the
default (HZ for frequency or S for time) is assumed.
SUBSYSTEM COMMANDS FORM DESCRIPTION
SENSe[1|2]:CORRection:COLLect
:MP:LOAD
<STAN1|STAN2|…|STAN12>
command
only
Measures a load on the port selected
during a test set calibration.
2
SENSe[1|2]:CORRection:COLLect
:MP:THRU
<STAN1|STAN2|…|STAN6>
command
only
Measures a thru on the port selected
during a test set calibration.
2
SENSe[1|2]:CORRection:COLLect
:PORTS <2|4|6|8|10|12>
NR1 Selects the number of ports to
perform a test set calibration on.
1
SENSe[1|2]:CORRection:COLLect
:SAVE
command
only
Completes and saves current
calibration.
SENSe[1|2]:CORRection:COLLect
:VERify:REFLection
<STAN1|STAN2|…|STAN12>
command
only
Measures a calibration standard
during a cal check procedure for
reflection measurements.
SENSe[1|2]:CORRection:COLLect
:VERify:TRANsmission
<STAN1|STAN2|…|STAN12>;
command
only
Measures a calibration standard
during a cal check procedure for
transmission measurements.
SENSe[1|2]:CORRection:CSET
[:SELect] DEFault
command
only
Restores the "factory" default
calibration for the current
measurement and channel.
SENSe[1|2]:CORRection:CSET
[:SELect]?
query only
CHAR
Queries the current calibration type
— returns DEF (factory default),
FULL (full band) or USER (user
defined).
SENSe[1|2]:CORRection:EDELay
:TIME <num>
2
NR3 Specifies the electrical delay in
seconds.
SENSe[1|2]:CORRection:EXTensi
on:REFLection[:TIME]<num>
2
NR3 Specifies the port extension at the
reflection port, in seconds.