11-52 Programmer’s Guide
SCPI Command Summary
SCPI Device Command Summary
Table 11-19 SENSe (15 of 16)
1. Numeric parameters may include an appropriate suffix; if no suffix is included, the
default (HZ for frequency or S for time) is assumed.
2. For use with structural return loss measurements on analyzers with Option 100
only.
3. Binary parameters accept the values of 1 (on) and 0 (off) in addition to ON and OFF.
SUBSYSTEM COMMANDS FORM DESCRIPTION
SENSe[1|2]:FUNCtion:SRL
:<num>,<num>
1
NR1,NR1 Specifies that the receiver will
measure the ratio of the power (SRL)
into the specified measurement
channel
2
.
For 8712ET/8714ET models—choose
1,0 (Forward A/R).
For 8712ES/8714ES models—choose
1,0 (Forward A/R), or 22, 20 (Reverse
B/R).
SENSe[1|2]:FUNCtion:SRL
:IMPedance <num>
1
NR2
Sets the cable impedance.
2
SENSe[1|2]:FUNCtion:SRL
:MODE <char>
CHAR Sets the auto z function to AUTO or
MANual.
2
SENSe[1|2]:FUNCtion:SRL
:SCAN[:IMMediate]
command
only
Starts a cable scan.
2
SENSe[1|2]:ROSCillator
:SOURce <char>
CHAR Specifies the source of the reference
oscillator — select INTernal or
EXTernal.
SENSe[1|2]:STATe <ON|OFF>
3
NR1 Turns the specified channel on/off .