1-148 Chapter1
Alphabetical Command Reference
NOOP
NOOP
Syntax
NOOP;
Description
Front Panel Equivalents
GPIB only: no front panel equivalent.
NUMG
Syntax
NUMG<num>;
Description
Front Panel Equivalents
Command Description Range Query Response
NOOP Creates a cycle that has no operation.
OPC-compatible.
N/A N/A
Command Description Range Query Response
NUMG Activates the indicated number of groups
of sweeps. A group is whatever is needed
to update the current parameter once.
This function restarts averaging if it is
enabled. OPC-compatible.
integers 1–999 N/A
Command Hardkey Softkey
NUMG
Sweep Setup
NUMBER of GROUPS