Chapter 7 7-133
Programming Examples
Limit Line and Data Point Special Functions
Example Program of OUTPSEGAF Using BASIC
The following program is not included on the “Programming Examples” CD-ROM:
10 OUTPUT 716; "outpsegaf;"
20 ENTER 716; Numsegs
30 PRINT "Receiving status for"; Numsegs; "segments."
40 IF Numsegs>0 THEN
50 FOR I=1 TO Numsegs
60 ENTER 716; Segnum, Pf
70 PRINT USING "DD, 2X, 8A"; Segnum, Pf
80 NEXT I
The example program shows how the OUTPSEGAF command can be used to request the
number of active segments and their status. Notice that each segment result must use a
new enter command as a line feed terminates each segment data result.
Output Minimum and Maximum Point Per Limit Segment
The command MINMAX<ON|OFF> toggles a feature which records the minimum and
maximum data points in all active limit segments. Note that limit testing need not be
turned on.
The command OUTPSEGM will report the min/max data for the segment previously selected
by SELSEG. The data is returned in a comma delimited string with the segment number,
minimum point stimulus, minimum trace value, maximum point stimulus and maximum
trace value.
Under the following conditions, OUTPSEGM will issue the following errors:
If the min/max testing is off: “30: Requested Data Not Currently Available. To clear the
error message, turn the min/max testing on.
If the limit table is empty: “204: Limit Table Empty” (this is a new message). To clear
the error message, enter a new limit table.
When the above error conditions occur, there is no data to report, thus no output is
generated.
If the selected segment has no associated limit, the NO_DATA string is generated,
which reports a stimulus value of 0 and a data value of 1000.

Table of Contents

Other manuals for Agilent Technologies 8720E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8720E Series and is the answer not in the manual?

Agilent Technologies 8720E Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8720E Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals