EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>8720E Series

Agilent Technologies 8720E Series Programmer's Guide

Agilent Technologies 8720E Series
541 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #302 background imageLoading...
Page #302 background image
2-10 Chapter2
Introduction to Instrument Control
Instrument Control using HP BASIC
System Setup and GPIB Verification
Figure 2-4 The Network Analyzer System with Controller
1. Connect the analyzer to the computer with an GPIB cable as shown in Figure 2-4.
2. Switch on the computer, and launch HP BASIC or HP BASIC for Windows.
3. Switch on the analyzer.
a. To verify the analyzer's address, press:
The analyzer has only one GPIB interface, though it occupies two addresses: one for
the instrument and one for the display. The display address is equal to the
instrument address with the least-significant bit incremented. The display address
is automatically set each time the instrument address is set.
The default analyzer addresses are:
16 for the instrument
17 for the display
CAUTION Other devices connected to the bus cannot occupy the same address as the
analyzer.
The analyzer displays the instrument's address in the upper right section of the
display. If the address is not 16, return the address to its default setting (16) by
pressing:
Local
SET ADDRESSES
ADDRESS: 87xx
16 x1 Preset

Table of Contents

Other manuals for Agilent Technologies 8720E Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8720E Series and is the answer not in the manual?

Agilent Technologies 8720E Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8720E Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals