System-Related Operations
Use the knob and arrows to select .sta file, and then press Select.
l You can configure instrument to power-down state from location 0 on power up. The factory default is to recall fac-
tory default state at power-on.
l Front Panel:
l SCPI: MEMory:STATe:RECall:AUTO
Error Conditions
Up to 20 command syntax or hardware errors can be stored in the instrument's interface-specific error queues. See
SCPI Error Messages for more information.
l Front Panel:
l SCPI: SYSTem:ERRor?
Beeper Control
The instrument normally beeps when an error is generated from the front-panel or remote interface. You may disable
the beeper, but not the click generated by pressing a front-panel key or turning the knob.
l This setting is non-volatile; it will not be changed by power cycling or *RST.
l Front Panel
l SCPI:
SYSTem:BEEPer:STATe {OFF|ON}
Agilent 33500 Series Operating and Service Guide 115