260 Chapter 8
Bluetooth Signals
Using Clock/Gate Delay
Using Clock/Gate Delay
The steps in this procedure build upon the previous procedure.
This function is available only when the payload data is continuous PN9 and is intended to be used during
bit error rate (BER) testing.
1. Press
Packet (DH1) > Payload Data > Continuous PN9 > Return.
This activates the configuration that generates clock and gate signals relative to the Bluetooth signal.
2. Press
Clock/Gate Delay > 4 > Symbols.
The clock and gate will be delayed 4 symbols to synchronize with the demodulated data signal from the
device under test (DUT) at the input of the BER analyzer.