358 Chapter 11
GPS Modulation
Real Time GPS
Figure 11-6 Real Time GPS Setup with Internal Clock
Configuring the External Reference Clock
1. Access the real-time GPS personality (Mode > More(1 of 2) > GPS > Real Time GPS).
2. Press
More (1 of 2) > GPS Ref Clk Ext Int to Ext.
3. Press
GPS Ref (f0) > 11.03 > kcps.
4. Connect the external reference clock source to the DATA CLOCK INPUT connector.
The maximum clock rate for this input connector is 50 MHz with a voltage range of
0.5 to 5.5 V.
5. Set the external reference clock source to the chip rate value entered in Step 3.
The maximum data rate for this input connector is 50 Mb/s with a voltage range of
0.5 to 5.5 V.
NOTE The chip rate of the external source must match the chip rate value set using the GPS Ref (f0)
softkey.
Figure 11-7 on page 359 shows what the signal generator display should look like after all the steps have
been completed. The CA Chip Rate, P Chip Rate, GPS Ref (f0), and GPS Ref Clk fields in the
status area of the display show the new parameters you set in this procedure. Notice the settings Satellite
ID, Doppler Shift, Data, and Data Mode, have not changed from the previous procedure, “Setting Up
the Real Time GPS Signal” on page 357.

Table of Contents

Other manuals for Agilent Technologies E4428C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4428C and is the answer not in the manual?

Agilent Technologies E4428C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4428C
CategoryInverter
LanguageEnglish

Related product manuals