Chapter 3 57
Basic Operation
Configuring the RF Output
9. Highlight the frequency item for point 8, then press Insert Item.
Pressing
Insert Item shifts frequency values down one row, beginning at point 8. Note that the original
frequency values for both points 8 and 9 shift down one row, creating an entry for point 10 that contains
only a frequency value (the power and dwell time items do not shift down).
The frequency for point 8 is still active.
10. Press
590 > MHz.
11. Press
Insert Item > -2.5 > dBm.
This inserts a new power value at point 8 and shifts down the original power values for points 8 and 9 by
one row.
12. Highlight the dwell time for point 9, then press
Insert Item.
A duplicate of the highlighted dwell time is inserted for point 9, shifting the existing value down to
complete the entry for point 10.
Activating List Sweep for a Single Sweep
1. Press
Return > Sweep > Freq & Ampl
This turns the sweep on again. No errors should occur if all parameters for every point have been defined
in the previous editing process.
2. Press
Single Sweep.
The signal generator will single sweep the points in your list. The SWEEP annunciator activates during
the sweep.
3. Press
More (1 of 2) > Sweep Trigger > Trigger Key.
This sets the sweep trigger to occur when you press the
Trigge r hardkey.
4. Press
More (2 of 2) > Single Sweep.
This arms the sweep. The ARMED annunciator is activated.
5. Press the
Trigge r hardkey.
The signal generator will single sweep the points in your list and the SWEEP annunciator will be
activated during the sweep.