Assembly Replacement
Inner Instrument Cover
2-6
Inner Instrument Cover
Tools Required
• T-10 driver
Removal Procedure
Refer to Figure 2-2 for this procedure.
1. Disconnect the power cord.
2. Remove the outer-cover from the signal generator. Refer to "Outer Instrument Cover" on page 2-4.
3. Place the signal generator flat and upright with the front panel facing you.
4. Using the T-10 driver, remove the twelve screws (1) from the inner cover (2).
5. Remove the inner cover.
Replacement Procedure
• Reverse the order of the removal procedures.
• Torque all T-10 screws to 9 in-lbs.