Assembly Replacement
A3 Power Switch
2-16
A3 Power Switch
Tools Required
• T-10 driver
Removal Procedure
Refer to Figure 2-7 for this procedure.
1. Disconnect the power cord.
2. Remove the front panel. Refer to "Front Panel" on page 2-8.
3. Disconnect the W8 ribbon cable from the A3 Power Switch by lifting the white tabs on the connector.
4. Using the T-10 driver, remove the four screws (12) from the A3 Power Switch.
5. Remove the A3 Power Switch.
Replacement Procedure
• Reverse the order of the removal procedure.
• Torque all T-10 screws to 9 in-lbs.
• Perform the post-repair adjustments and performance tests that pertain to this removal procedure.